ELECTRICAL INSULATION TESTS AND MEASUREMENTS
Test type Insulation strength measured at warm before connection to test bench
Test procedure No   Cryomagnet id. HCLBBL_000-IN002049 Date of test 30/04/2004
Magnet Temperature 297.00 Kelvin Test configuration Main and Auxiliary magnets Test operator M. Makhlouf
Bench Name Rack HV barcode      
Polarity + Polarity - Ambient
Temp.
[K]
Ambient
Humidity
[%]
Voltage
[V]
Duration
[mn:s]
Final Current
[A]
Resistance
[W]
Capacitance
[F]
Dipole-Quench Heaters Ground 297 33 613 02:00 2.907E-8 2.107E+10 4.30E-7
                 
Dipole Quench Heaters - Ground 297 33 537 02:00 3.501E-8 1.533E+10 5.30E-7
                 
Dipole Ground 297 33 613 02:00 1.284E-8 4.773E+10 3.50E-7
                 
Dipole All Quench Heaters 297 33 537 02:00 2.517E-8 2.131E+10 4.10E-7
                 
All Quench Heaters Ground 297 33 613 02:00 2.845E-8 2.154E+10 3.30E-7
                 
Dipole MCS Correctors 297 33 256 01:01 4.851E-9 5.286E+10 0.00E+0
                 
MCS Correctors Ground 297 33 256 01:01 5.382E-9 4.764E+10 0.00E+0
                 
External and Internal Passive Busbar Ground 297 33 613 02:00 6.109E-9 1.003E+11 2.00E-8
                 
External Passive Busbar Internal Passive Busbar 297 33 613 02:00 2.753E-9 2.225E+11 0.00E+0
Accuracy of measurements ± 1K ± 2% ± 2% ± 1V   ± 5% ± 0.2nA ± 5% ± 5% ± 10nF
  Automatic HV analysis: OK